The Model 2052 is a portable, digital display conductivity instrument with three modes of operation: two for measurement of solution conductivity with a full scale from 0 to 200 millimhos, and one for self-check. The self-check mode provides verification of proper operation, the ability to adjust for cell constant variations and Total Dissolved Solids (TDS) scale factor. The conductivity and TDS measurement data is displayed on a 31/2; digit, LCD display with a floating decimal.
The function switch selects power off, Automatic Temperature Compensation (ATC) on, off, or self-check modes. ATC is set at 2% per °C at 25°C with reference at 25°C. Self-check (cell constant adjust) span is from 40% to 110% of nominal to allow cell constant correction and TDS scale factors to be easily set. Cell drive is a low amplitude sine wave, rectification is synchronous.
Ordering Information: Meter comes with 9V alkaline battery, 473mL (1pt.) 718 micromho standard solution and manual. Meter does not include probe. Lab kit includes two-piece stand with pole, cell holder, and AC adapter. Cell connector comes with 1m (3'3") cable.
|